Search - Content

H301-T-UNIT-BL-PLUS

H301-T-UNIT-BL-PLUS is the most advanced solution for Temperature Control. It is compatible with Digital Gas Controllers and with Active Humidity Controller to ensure best environmental control for live cell imaging. 

H301-T-UNIT-BL-PLUS has five independent temperature channels: chamber base, chamber lid, objective heater, sample temperature sensor, room temperature sensor.

  • Temperature Range: from 3°C above ambient to 60°C
  • Temperature Resolution: 0.1°C
  • Temperature Accuracy: ± 0.1°C in sample feedback mode, ± 0.3°C in chamber feedback mode
  • Controls up to # 5 Temperature sensors:
    • Chamber Base
    • Chamber Lid
    • Sample T Sensor
    • Room T Sensor
    • Objective Heater T sensor
  • Compatible with:
    • Objective Heater
    • Active Humidity Controller
    • Digital Gas Controllers
    • Smart Box data logger
  • SDK available for integration in third party software
  • Drivers available for: NIS Elements, Metamorph, Micormanager
 

Specimen-vs Chamber-feedback.jpg

download Data Sheet for complete description of the experimental methodology

H401-T-DUAL

H401-T-DUAL is a 2-channel button display T controller which can be employed to control the stage top chambers, the heated plates and / or the objective heater. It is not compatible with gas or humidity controllers

FEATURES

  • Temperature Range: from 3°C above ambient to 60°C
  • T resolution: 0.1°C
  • Chamber feedback operation
  • Data logging via USB and DATA LOG software
  • Button Display user interface

H401-T-DUAL for Stage Top Chambers

  • One T channel controls chamber base and one T channel controls the lid
  • Temperature Accuracy: ± 0.3°C if room remains within ± 1°C
  • Feedback mode: chamber feedback
  • YES: it is a good solution if Temperature control of the incubating chamber is the only requirement
  • NO: it is not a good solution if Gas and Humidity control is also required

H401-T-DUAL + warm plates

  • YES: it is a good solution in combination with two devices. See also H401-T-DUAL-BL
  • NO: it is not a good solution in combination with only one device. In this case, see H401-T-SINGLE and H401-T-SINGLE-BL

 H401-T-DUAL + Objective Heater

  • YES: it is a good solution to control the objective heater + one warm plate. See also H401-T-DUAL-BL
  • NO: it is not a good solution in combination with the objective heater only. In this case, see H401-T-SINGLE and H401-T-SINGLE-BL